"USB OSCILLOSCOPE" PROGRAM
USER MANUAL
CONTENT
1 PURPOSE
3
2 COMMON KNOWLEDGES
3
2.1 General functions
3
2.2 Minimal resource requirements
3
3 HOW TO WORK WITH THE PROGRAM
4
3.1 Program general controls description
4
3.1.1 Menu
4
3.1.2 Toolbar
4
3.1.3 Control panel
4
3.1.4 Popup panel
5
3.1.5 Status bar
7
3.2 Start oscilloscope
7
3.3 Analogue oscilloscope mode
8
3.3.1 Sampling frequency control (time/div)
9
3.3.2 Horizontal zoom control
9
3.3.3 Channel parameters control
10
3.3.4 Channel parameters control panel
10
3.3.5 Synchronization parameters control
11
3.3.6 Synchronization parameters control panel
12
3.3.7 Horizontal scroll control
14
3.3.8 Active channels view parameters auto-adjusting
14
3.3.9 Vertical composition of channels
14
3.3.10 Analogue inputs calibration
15
3.4 Specter analyser mode
17
3.4.1 Specter analyzer parameters control
18
3.4.2 Specter parameters control panel
18
3.5 Logic analyser mode
19
3.5.1 Logic analyzer synchronization parameters control
20
3.5.2 logic analyzer synchronization control panel
20
3.6 Data storing
21
3.6.1 Storage file location definition
21
3.6.2 Maximal storage time limitation
21
3.6.3 Storage time interval definition
22
3.6.4 Storing of predefined time interval
22
3.6.5 Definition of markers during storage
22
3.7 Oscillogram view mode
23
3.7.1 Oscillogram saving
23
3.7.2 Searching in oscillogram file
24
3.7.3 Search parameters control panel
25
3.7.4 Selection of oscillogram fragment
25
3.7.5 Horizontal auto-zooming
26
1
3.7.6 Saving of oscillogram fragment
26
3.8 Measurement tools
26
3.8.1 Measurement panel
27
3.8.2 Measurement markers
27
3.8.3 Signal phase and off-duty factor measuring
28
3.8.4 Signal values recalculation and units types
29
3.8.5 User units types editing
29
3.9 User preset files
31
3.9.1 Saving of current settings in UP file
31
3.9.2 Loading or deleting UP files
31
3.10 Marking of oscillogram
32
3.10.1 Bookmark set or delete functions
32
3.10.2 Bookmarks list window
32
3.10.3 Navigation through oscillogram with the usage of bookmarks and search by
comment
32
3.10.4 Bookmark play function
33
3.11 Additional tools
33
3.11.1 Using of level/range markers
33
3.11.2 Oscillogram file description
34
3.11.3 View options
34
3.12 Screen memory
34
3.12.1 Current screen remember function
34
3.12.2 Saving, loading and clearing of the screen memory
34
3.12.3 Changing of screen memory transparency
34
3.13 Oscillogram printing
35
3.13.1 Print preview window
36
3.14 Saving of oscillogram fragment in image file
36
3.14.1 Selecting of oscillogram area
36
3.14.2 Saving oscillogram fragment as a picture
37
3.15 Plug-ins
37
3.15.1 Using of Plug-ins
37
3.16 Oscillogram data automatic analysis
39
3.16.1 Using of analyser scripts
39
3.16.2 Analyzer panel usage
39
3.16.3 Graphical results printing
40
3.16.4 Saving of graphic in a picture file
41
3.17 Emulation mode
42
3.17.1 Running of emulation
42
3.17.2 Emulation activation
42
2
1 PURPOSE
The main purpose of the "USB Oscilloscope" program (hereinafter referred to as
Program) is monitoring, storage, analysis and measuring parameters of input
signals, digitized by the USB-oscilloscope device
(hereinafter referred to as
Device).
2 COMMON KNOWLEDGES
This program works under Windows 2000/XP, Windows Vista and Windows 7
operation systems, based on x86 or AMD64 CPU family PC.
2.1 General functions
The program is used for simultaneous visualization, analysis and recording of the
signals, digitized by the Device in real time. Data is stored as a continual time
fragment. The fragment size is limited by a maximum file size, and is equal to 16 T
bytes for NTFS or 4 G bytes for FAT32 file systems.
Program is equipped with tools for automatic analysis of recorded signals on the
basis of the script core. The results of its work are text and graphical reports,
marking and commenting of oscillogram specific points. There is a possibility to
export analysis results to a text file.
Usage of plug-ins allows performing specific signals visualization and analysis
with getting the graphics of the researched parameters changes in real time.
The program is equipped with the functions of printing and saving in a graphical
format of selected oscillogram fragments and analysis results.
Emulation mode helps to visualize a signal dynamic, using the earlier recorded
oscillogram file.
2.2 Minimal resource requirements
-
CPU not less than Pentium III 1 GHz;
-
graphic adaptor not less than SVGA 800x600;
-
RAM not less than 1 G byte.
Attention! Be sure that the Device is connected to the USB-port of your PC
before starting the real time signal monitoring. Please check the Device drivers to
be properly installed.
3
3 HOW TO WORK WITH THE PROGRAM
3.1 Program general controls description
The Program main window (fig. 1) has the following control items:
1 - menu;
2 - popup menu;
3 - tool bar;
4 - control panel;
5 - status bar.
fig. 1. Program main window.
3.1.1 Menu (fig.1 [1]) contains elements, which are grouped according to their
functionality into popup menu
(fig.1
[2]). To call the required function it is
necessary to activate the corresponding item of Program’s menu and then to
select the needed item in the popup menu. This operation can be performed with
the help of the mouse cursor or by using the keyboard.
3.1.2 Toolbar
(fig.1
[3]) contains control items for calling some Program
functions. These items are presented as buttons with icons. Activation of the
toolbar functions is carried out with the help of the Mouse cursor.
3.1.3 Control panel (fig.1 [4]) contains control items, placed in the direction from
the top to the bottom. These items display the function name, one or a group of the
controlled parameters and control buttons, placed in the right part of the item.
4
Control panel has the following types of items:
- simple button;
- popup panel or list control;
- popup panel control with fast function;
- parameter increase/decrease control;
- select/increase/decrease parameter control.
The choice of the active item of the control panel is performed by the mouse
cursor or keyboard up/down arrow keys. The function of items like button, popup
panel or list control is activated by clicking the left mouse button or the keyboard
right arrow, “Space” or “Enter” keys.
The similar functionality has popup panel control with fast function item type.
But in this case pressing of the button with the help of the mouse cursor is used
to display popup panel, and the button can be used for the given item fast
function execution. To call the fast function for the current item type with the help
of the keyboard, it is necessary to press and hold the left arrow button at least for
0.5 seconds.
To change the parameter value for increase/decrease control item type, press
/
item buttons with the help of the mouse cursor. The same functions have
keyboard left or right arrow keys.
Similar functionality has select/increase/decrease control item type. In addition
the given item part has the central button. Pressing of this button with the help of
the mouse cursor allows selecting one of the parameter values from the dropped
list. The given function is duplicated by the “Space” or “Enter” keyboard keys.
3.1.4 Popup panel (fig. 2) is an extension of the control panel and contains
control elements, grouped separately.
fig. 2. Popup panel.
Use mouse cursor or keyboard left/right arrow keys to change the active item of
the popup panel. The panel active item has the control part, placed in its right side
([1] fig.1). If it is not possible to activate items, they are temporary disabled. The
last item of this panel is the close button. It can be used to close popup panel.
Alternatively popup panel is closed by pressing the keyboard “Esc” key or by
moving the active item out of leftmost item with the help of the left arrow key.
5
Popup panel has the following items types:
- button;
- switch;
- submenu;
- select control;
- increase/decrease control;
- select/increase/decrease control;
- text enter/select.
Press item body with the mouse cursor or use keyboard down arrow or “Enter”
keys to execute button, switch and submenu items types function.
Press button in the right part of select control item type with the mouse cursor
or down arrow or “Enter” keyboard keys to show options of the dropping list. Then
use the mouse cursor or keyboard up/down arrow keys to activate the required
item of this list. Press the left mouse button or keyboard right arrow or “Enter” keys
to accept the selected item. Press keyboard left arrow or “Esc” keys to cancel any
selection.
Press
/ buttons with the help of the mouse cursor or keyboard up/down arrow
keys to change the value for increase/decrease item type.
Similar functionality has select/increase/decrease item type. In addition, the
value for this item type can be selected from the dropped list. This list is shown
when the middle part of item body is pressed with the mouse cursor or keyboard
“Enter” key can be used instead.
The text enter/select item type helps to enter/select text or numerical values.
This type of item is mainly controlled with the help of the keyboard. The keyboard
left/right arrow and “Home”/“End” keys are used to move the caret, “Backspace”
and “Del” keys help to edit the text. This item type supports standard clipboard
functions.
General items of control and popup panels are controlled by four
keyboard keys: arrow left, right, up, down.
Keyboard up / down arrow keys are used to select the active item of the control
panel and left / right arrow keys help to control it. And vice versa: keyboard left /
right keys are used to select the active item of the popup panel and up / down
arrow keys help to control it. The active item of the dropped list can be changed
with keyboard up / down arrow keys, right arrow key is used to select the item, left
arrow key to cancel selection. In addition increase/decrease and switch type items
can be controlled with mouse wheel rotation.
Control panel contains general set of control elements that control oscilloscope
and view functions. Therefore it is possible to control Program’s general functions
with four keyboard keys.
6
3.1.5 Status bar (fig.1 [5]) shows the current item information - its function
description and hot keys combination for the given item.
3.2 Start oscilloscope
Use the “Start device” item of the control panel to initiate the oscilloscope start
configuration, which can be done with the help of the start device control panel fig.
3.
fig. 3. Start device control panel.
Start device control panel items (fig. 3 from left to right):
- current device will be used as a data source
;
- current device mode
(depends on which device is selected) can be one of:
- for analogue oscilloscope mode;
- for logic analyser mode;
- number of input channels
- manual sampling mode .
Use the button to accept current device settings and start it. Depending on the
chosen device the given panel can have less items, described above or have
additional ones.
Alternative methods for device running are to use UP files (see 3.9.2) or plug-
ins (see 3.13.1) Program functionality.
7
3.3 Analogue oscilloscope mode
In the analogue oscilloscope mode the Program processes analogue signals,
digitized by the current Device. The program window in this mode is shown on fig.
4.
fig. 4. Program window in analogue oscilloscope mode
The Control panel has the following control items (fig. 4.):
[A] - stop device;
[B] - load user preset file;
[
] - switch on/of data recording;
[D] - stop/start frame (synchronization status);
[E] - sampling frequency control;
[F] - horizontal zoom control;
[G] - 1 - channel control;
[H] - synchronization parameters control;
[I] - horizontal scrolling control;
[J], [K] - markers A and B position control;
8
[L] - phase/off-duty factor measurement control.
Numbers on figure 4 show the following control items:
1 - pointer of channel 1 zero level vertical position;
2 - frame synchronization point horizontal offset;
3 - current synchronization level;
4,5 - measurement markers A and B accordingly;
6 - horizontal scroll bar control;
7 - vertical scroll bar control;
8 - measurement panel.
3.3.1 Sampling frequency control (time/div)
Sampling frequency is controlled by the item [E] fig. 4 of the control panel. Icon,
placed in the left part of this item, displays current sampling mode:
- frame;
- sub frame (stroboscopic);
- stream.
Devices, supporting different sampling modes, can change one of the above
mentioned sampling modes, depending which sampling frequency is currently
selected.
In the frame sampling mode data is transmitted from the Device as a small
fragment of the input signal(s), captured in accordance with the conditions of
synchronization frame by frame. Sub frame sampling mode is similar to frame one,
but in this mode each portion of Device data is the part of the whole frame. In the
stream sampling mode data is digitized continuously.
The current item dropping list displays a set of all possible for the current device
sampling frequencies and corresponding time/div values, according to the current
horizontal zoom value.
3.3.2 Horizontal zoom control
The value of the current horizontal zoom is controlled by the item [F] fig.4 of the
control panel. Zoom value defines a number of samples, used for single
oscillogram dot displaying. The item allows using the oversampling and in
combination with the data display mode allows displaying the signal as average or
peak values (see 3.3.5). The change of the zoom value gives the opportunity to
compress or stretch signal horizontally without changing the sampling frequency. A
zoom control item displays current zoom and corresponded time/div values
(fig. 4. [F]).
9
3.3.3 Channel parameters control
Channel parameters are controlled by the item [G] fig. 4 of the control panel.
Item body displays (from left to right):
- channel number
;
- analogue input is used by this channel
;
- vertical division value in terms of scale step;
- voltage range for analogue input is used for current channel.
When the signal inversion option is active, the vertical division value is
displayed with the negative sign.
The item activates channel parameters control panel. The fast function of the
given item sets the optimal vertical division value and centers the signal vertically.
3.3.4 Channel parameters control panel
Channel parameters control panel is shown on fig. 5.
fig. 5. The channel parameters control panel.
As it is shown on the figure above, the panel contains the following control items
(from left to right):
- used analogue input number
… , or
- when channel is temporary
switched off;
- combination of input voltage range and input divider for chosen
analogue
input;
- vertical division value control
;
- vertical offset value control in relative units
;
- vertical offset value reset
;
- compensation mode for signal offset:
- off compensation;
- normal compensation mode;
- manual compensation mode;
- signal offset compensation value in manual mode
;
- automatic calculation of the signal compensation value
;
- switch on
/ off
signal inversion;
- view mode: peak to peak or averaged
;
- switch on
/ off
channel subtraction;
- show
/ hide
channel value display in measurement panel;
- channel color select
;
- channel (input) name;
- show analogue input parameters adjusting panel .
10
The panel item, that defines the input voltage range, helps to select the
combination of the input range (if this possibility is supported by the device/input,
currently used) and input divider values, used for the current input. As it is shown
on figure 5 the input voltage range is ±6 volts and the input divider is 1:10, or in
other words it equals to ±60 volts.
Signal zero line offset depends from current compensation mode. When
compensation mode is switched off
,
- signal zero line defines by relative offset
only. Vertical relative offset defines signal offset from the center of the screen
area defined for current signal (see 3.3.9 vertical composition of channels) in per
cents of the screen height. In the rest of compensation modes, zero line position
defines by two values. First is vertical relative offset. Second services to
compensate constant offset of given signal. This value moves signal zero line so
that signal informative part still on same place during vertical division value
changing. Compensation mode and signal offset compensation value are
associated with analogue input is currently selected.
The channel subtraction function helps to show the differential signal for two
neighbor channels. This signal is calculated as a difference between the current
cannel - A and the next channel - B, by formula: A = A - B. In addition when
channel A has the inversion, the differential signal is also inverted, and when
channel B has the inversion this inversion is used before the channels subtraction.
3.3.5 Synchronization parameters control
Synchronization parameters are controlled by the item [H] fig.4 of the control
panel. This element displays current synchronization options (from left to right,
from the top to the bottom):
- channel is used as a synchronization source
;
- front:
- raise;
- fall;
- minimal synchronization interval
;
- synchronization mode:
- auto-level;
- normal;
- wait;
- single;
- switch off synchronization;
- current synchronization level .
This item activates the synchronization parameters control panel. Fast function
for this item helps to adjust synchronization level automatically.
11
3.3.6 Synchronization parameters control panel
Synchronization parameters control panel is shown on fig.6.
fig. 6. Synchronization parameters control panel.
As it is shown on figure above the
panel
contains
the
following
items
(from left to right):
- channel is used as synchronization source
- synchronization mode:
- auto-level;
- normal;
- wait;
- single;
- switch off synchronization;
- front:
- raise;
- fall;
- synchronization level
;
- relative synchronization level
;
- synchronization level auto-adjusting
;
- level increasing threshold
;
- number of points for signal averaging
;
- minimal synchronization interval mode:
- minimal interval for signal period;
- minimal interval for signal impulse;
- minimal synchronization interval value
;
- minimal synchronization interval reset
;
- synchronization point offset
;
- synchronization point offset reset
;
- synchronization delay .
A single oscilloscope channel is used as
a
synchronization
source.
Four
synchronization modes can be used for the synchronization, or the synchronization
can be switched off. The synchronization modes description is below.
Normal mode always displays a signal, whether the synchronization point
was found or not. If the synchronization point is not found during the
synchronization delay time interval, the Program displays the not synchronized
signal.
Auto-level mode is functionally similar to the normal mode. In this mode the
synchronization level is controlled automatically, that is useful for the investigation
of dynamically changing signals. The synchronization level in this mode is
12
calculated, depending on the signal amplitude parameters and current value of the
relative level, which is defined by the item - . For example when the relative level
equals to 50% it corresponds to the middle level from minimal and maximal signal
values. The 10% value differs on 10% from the minimal toward the maximal signal
level, and 90% is 10% less than the maximal signal level accordingly.
Wait mode in this mode screen data updates only when the synchronization
point is found. In other words if there is no synchronization the signal on the
screen is frozen until the next synchronization point is found.
Single mode is similar to the wait mode, screen data updates only when
synchronization point is found. Then screen data is unchanged until
synchronization will be restarted by “Frame start” button ([D] fig. 4) of the control
panel.
Temporary frame can be stopped or started again with the help of the “Stop
frame” button ([D] fig. 4).
Front and synchronization level defines general synchronization parameters.
These parameters are used for searching the synchronization point - the point
where the signal crosses the current level in a given front.
Synchronization level auto-adjusting is accessible in all synchronization
modes except auto-level mode. The button of synchronization panel is used to
run this function. During auto-level adjusting process, like in auto-level mode,
Program evaluates the synchronization level from parameters of input signal (see
description of auto-level mode).
Level increasing threshold is an additional selectivity parameter that
determines the minimal increasing voltage of the signal at the synchronization
point. Searching process continues, if the signal increasing voltage is less than the
defined threshold value. Increasing of this parameter allows to ignore flat signal
fronts and to synchronize on the steeper ones. The number of signal averaging
points is also an additional selective parameter. It defines the number of points,
used for signal averaging at the synchronizer input. Increasing of this parameter
helps to filter noisy signals that give more stable synchronization for flat front
signals. The given parameter is used in combination with the threshold value.
Thus, these selective parameters help to synchronize high frequency signals by
increasing the value of increasing threshold or low frequency by increasing the
number of signal averaging points.
The next parameter - minimal synchronization interval, helps to make an
additional selection in the synchronization points set. It defines the value of the
minimal time interval between two synchronization points. In other words, this
parameter helps to synchronize signals that have a set of seriated fronts, and
allows selecting the first front in series. The time interval inside series of fronts is
always less than the interval between two series. Thus, if the value of the minimal
synchronization interval is more than the time interval inside fronts series, the
synchronization point will be chosen at first front in series.
Synchronization point offset
defines offset of the synchronization point
relatively to synchro-frame center. The change of this parameter allows moving
synchronization point so that to place the signal fragment in the center of the
synchro-frame. By default this point is placed in the center of synchro-frame.
13
Synchronization delay determines the time interval for synchronization
searching process in normal and auto-level synchronization modes
(see
description for normal synchronization mode). If there is no synchronization,
increasing of this parameter decreases the screen update frequency.
The parameters of level increasing threshold and numbers of signal averaging
points can be inaccessible in the frame and sub frame sampling modes.
3.3.7 Horizontal scroll control
The horizontal scroll control is duplicated by item [I] fig.4 of the control panel.
This item displays the offset value of the leftmost point of the current screen in a
frame or an absolute position of the current screen in a file. The offset value is
displayed in terms of time or frequency for the spectrum analyzer mode. This
control item is supplied with dropping list that performs fast scrolling functionality
with 10% resolution.
3.3.8 Active channels view parameters auto-adjusting
The Oscillogram screen is spitted vertically to display all active signals.
Functionality of this function is similar to the channel view parameters adjusting,
executed by the channel fast function (see 3.3.4). But in this case the function
adjusts view parameters for all active channels simultaneously. Press the button
of the toolbar, menu option
“Operations\Arrange waves” or Ctrl+G hot key
combination to run this function.
3.3.9 Vertical composition of channels
The Program automatically divides the current height of the screen in such way
to provide displaying of all active signals in the increasing order of the channel
number. Thus each channel is displayed in the corresponding area. If necessary
the given order can be changed with two ways. The first one is to drag the pointer
of the channel zero level with the help of the mouse cursor with "Ctrl" key is held.
This operation is performed in the following order: the "Ctrl" keyboard key is
pressed and held, and then the channel zero level pointer is captured with the help
of the mouse cursor, by pressing the left mouse button. The screen displays the
window of the channels composition. The new order for the current channel is set
by the movement of the item with given channel number into the required position.
The screen displays at the same time current variant of the signals vertical
composition. So in such way it is possible to change the order of the signals
displaying or join current signal with other in the group. The operation is finished
with the left mouse button releasing. If necessary it is possible to cancel any
composition changes by pressing the "Esc" keyboard key. The second way to
change the channels composition is to use the “View/Channels composition” menu
option. The function of the given menu displays the described above channels
composition window. By analogy with the previously described mechanism, the
current channels composition can be changed by a simple dragging of the channel
number item with the help of the mouse cursor. Alternatively the channel
composition can be modified with the help of the keyboard. The channel number,
which composition is changed, is set by the channel number "1", "2", etc. key
14
pressing, the arrow up/down keys change the current composition. The new
composition is set by pressing the "Enter" key, the "Esc" key cancels any
composition changes.
3.3.10 Analogue inputs calibration
Analogue input calibration panel is used for the current Device analogue input
calibration process. The view of this panel is shown on fig. 7.
. 7. Analogue input calibration panel.
Panel contains the following control items (from left to right):
- volts on ADC step value
;
- reference voltage for volts on ADC step value calculation
;
- volts on ADC step value auto-adjusting
;
- zero offset value
;
- zero offset value auto-adjusting
;
- previous calibration values loading
;
- current calibration values saving
;
- current calibration values accepting .
Calibration can be done manually or automatically. In first case, volts on ADC
step and zero offset values can be set by using corresponding items of the
calibration panel. In automatic mode the Program calculates these parameters in
accordance with the input signal level.
Automatic calibration is performed in the following order:
First of all the auto-adjusting of zero offset value is done. It is necessary to
disconnect any input signals from the current input, and it is desirable to connect
the input with the measuring ground. This function is run by the button
of the
calibration panel. Then the auto-adjusting of volts on ADC step value is
performed. The reference voltage source must be connected to the calibrated
input, and its voltage value must be entered or selected in the reference voltage
value item
of the calibration panel. The function is run by the button
of the
calibration panel.
Previous calibration values for current analogue input can be reloaded by the
button
of the calibration panel. Use the same button with the combination of
“Ctrl” keyboard key to load device default values for the current analogue input.
Use button of the calibration panel to save current calibration values. When the
button of the calibration panel is used without saving, any changes will be used
until the Device is stopped. When calibration panel is closed without saving, all
changes are discarded.
The calibration of the analog input must be performed with the input divider
value equal to 1:1 (x1). In other case the current input divider correction will be
done for the current analogue input. This type of calibration can be used to
15
compensate input divider inaccuracy. It is strongly recommended to do input
divider calibration only on calibrated analogue input. Only volts on ADC step value
adjusting should be performed for input divider calibration. The result of this
calibration is saved as a correction value for combination of current input divider
and analogue input number.
In addition, Program has a set of functions, grouped in popup menu
“Control/Analog inputs parameters”. The first function of this group is
“Load
defaults”. It helps to restore default values for all analogue inputs of the current
Device. The next one is “Save current”. It is used for saving calibration values for
all analogue inputs of the current Device. The “Adjust zero offsets” menu function
helps to perform zero offset auto-adjusting for all active inputs of the current
Device. Like in case with a single input zero offset adjusting this function requires
as minimum disconnection of all active inputs from any signal sources.
The Current Device built in input dividers values can be changed with “Set built-
in divider” menu item function.
16
3.4 Specter analyser mode
This mode is used to display specters of input analogue signals. The Program
window in this mode is shown on fig. 8.
Fig. 8. Program window in specter mode
As it is shown on figure above, control panel in this mode is similar to the
analog oscilloscope mode. Zoom control item is replaced with the specter analyzer
parameters control item [1] fig.6.
Analogue oscilloscope/specter analyser mode switching is performed with
the toolbar button
,
“View/View as specter” menu item or Ctrl+W hot key
combination. The synchronization control in this mode is unavailable. It is
recommended to use analogue oscilloscope mode to set up synchronization
parameters, and then switch Program in specter analyser mode. Horizontal scroll
and marker A and B control items, display its values in terms of frequency. This is
because horizontal screen axis in this mode represents frequency space.
17
3.4.1 Specter analyzer parameters control
Specter analyzer parameters are controlled by specter parameters control item
[1] fig.8. This item displays the following items (from left to right):
- type of vertical scale is used:
- linear;
- logarithmical;
- number of specter lines are currently used
;
- measurement scale step value .
This item activates specter parameters control panel.
3.4.2 Specter parameters control panel
Specter parameters control panel is shown on fig. 9.
Fig. 9. Specter parameters control panel.
Panel contains the following control items (from left to right):
- current number of specter lines
;
- type of vertical scale:
- linear;
- logarithmical;
The number of specter lines defines the number of dots in spectrogram.
Increasing of the given parameter decreases the frequency step between two
spectrogram dots, making it in such a way more detailed.
When the linear scale type is chosen, each dot of spectrogram represents
amplitude value of current frequency component in the input signal in terms of
volts. If the logarithmical scale type is chosen all spectrogram values are
presented in db (decibel). In this case 0 db is assumed to the signal, which
amplitude equals to the current input range (for instance, if the input range is
±15 V it is 30 V). Vertical scale step in this case is also defined in terms of db
units.
18
3.5 Logic analyser mode
In this mode Program processes two levels logical signals, digitized by the
Device. The view of the Program window in this mode is shown on fig.10.
fig. 10. Program window in specter analyser mode
As it is shown on fig. 10 control panel has similar control items with analogue
oscilloscope mode.
19
3.5.1 Logic analyzer synchronization parameters control
Synchronization parameters are controlled by item [1] fig. 10 of the control panel.
This element displays the following synchronization options (from the top to the
bottom, from left to right):
- synchronization mask;
- synchronization mode:
- normal;
- wait;
- single;
- switch off synchronization;
- minimal synchronization interval value .
Synchronization mask is a condition or a combination of synchronization
conditions for single or multiple channels. This element activates logic analyzer
synchronization control panel.
3.5.2 logic analyzer synchronization control panel
Logic analyzer synchronization control panel is shown on fig. 11.
fig. 11. logic analyzer synchronization control panel.
As it is shown on fig. 11, the panel contains the following items (from left to
right):
- synchronization mode:
- normal;
- wait;
- single;
- switch off synchronization;
- synchronization condition control items for all active channels, with possible
conditions:
- raise;
- fall;
- low level;
- high level;
- not important;
- minimal synchronization interval value
;
- minimal interval value reset
;
- synchronization point offset
;
- reset of synchronization point offset
;
- synchronization delay value
;
20
Synchronization condition mask is used to synchronize logic analyser signals.
The mask defines any combination of the front and levels. Front condition can be
set only for a single channel. Therefore the condition mask shown on fig. 11 is as
follows: rising front on channel 1, low level on channel 2, high level on channel 3,
the condition of channel 4 is not important. Minimal interval value defines minimal
time interval between two synchronization conditions for the defined mask. The
rest of items have identical functions to analog oscilloscope mode, see 3.3.6.
3.6 Data storing
Data storing function is available in all Program modes and does not depend on
current signals visualization mode. The maximal file size is limited by the used file
system. For instance FAT32 maximal file size equals to 4 G bytes, and for NTFS it
is 16 T bytes. The maximum recording time is directly proportional to the available
disk space, and inversely proportional to the sampling frequency. In other words,
for the same resulting file size with increasing sampling frequency, the maximum
storage time reduces and vice versa. The right part of the status bar displays the
maximal storage time that is available for the current Program mode in the
following format: H:M:S, where H - hours, M - minutes and S - seconds.
Data storage starts or stops by “Store data” item of the control panel. In stream
sampling mode all data is stored as a continues time fragment. Data in frame
mode is stored as separate frames. Therefore in the first case, the storage control
item displays the total time of the stored data fragment, in the second one - the
number of stored data frames.
fig. 12. Storage control item during storage in the stream sampling mode.
The storage time is displayed in format H:M:S:tS see fig. 12, where H - hours, M
- minutes, S - seconds, tS - tens of seconds. For the frame sampling mode, the
data is stored only after data frame updating. The “Stop frame” function has no
affect on data storage process.
3.6.1 Storage file location definition
By default data is stored into the Program root folder. This location can be
changed with the help of “Control/Data storage Options/Define storage location”
menu option. This option can be useful when there is no wish to store data on the
system disk, where the available disc space is limited. Storage location changes
will be available only after program restarting.
3.6.2 Maximal storage time limitation
Storage time can be limited with the help of
“Control/Data storage
Options/Enable time limitation” menu option. Storing will be automatically stopped
after storage of the defined time interval see 3.6.3.
21
3.6.3 Storage time interval definition
Value of storage time interval is used for storage limitation or circling see 3.6.2
or 3.6.4. Use “Control/Data storage Options/Storage time interval” menu option to
change this value.
3.6.4 Storing of predefined time interval
This storage option is activated by “Control/Data storage Options/Cycling of
data storage” menu option. When the option is active data storage proceeds in
such a way: first of all the defined time interval is stored, and then recording starts
circling. Therefore when data storage is stopped, the file contains predefined time
interval (see 3.6.3), which ends at the moment of storage stopping. In this case the
storage control item displays firstly the current storage time of the accumulation
process, and then the “R” flashing symbol for the circling stage.
3.6.5 Definition of markers during storage
Program allows setting up to
50 markers during data storage to define
characteristic parts of oscillogram. It is necessary to pay attention to the fact that
while “Cycling of data storage” option is active, all markers, indicating the fragment
of the oscillogram, rewritten with new data portion, are destroyed.
22
3.7 Oscillogram view mode
This mode is activated each time after data storage ending, or binary
oscillogram file opening (file with *.mwf - extension). This mode Program window is
shown on fig. 13.
fig. 13. Program window in oscillogram view mode.
General control items in this mode are identical to the items described above.
Synchronization control item in this mode is replaced by the search control item.
3.7.1 Oscillogram saving
All changes of view, channels settings, bookmark, level markers and etc., made
by the user, can be saved in the oscillogram file. Toolbar button
, “File/File Save”
menu option or Ctrl+S hot keys combination can be used to save the oscillogram
file. The Program is additionally supplied with “File/File Save as..” menu option that
can be useful when the file name/location or data compression option should be
changed. The data compression option can essentially reduce oscillogram file
size, but it can increase file opening time during decompression. Definition of the
23
oscillogram file name, location and compression options is performed with the help
of oscillogram saving dialog, shown on fig. 14.
fig. 14. Oscillogram file save dialog.
As it is shown on fig. 14, data compression options are in the bottom side of the
dialog. The “Save compressed” option allows to use or cancel data compression
during oscillogram data saving. The “Delta compression” option controls additional
compression option, used for analogue data compression. The
“Compression/speed” option helps to select the compression rate, which will be
used during data compression.
3.7.2 Searching in oscillogram file
This function is used for searching inside analogue or logic analyser oscillogram
files. Searching function is unavailable in the specter analyzer mode. Current
settings of the searching function parameters are displayed by the corresponding
item the control panel fig. 15 and 16.
fig. 15. View of the searching control item for analogue data.
This item displays the following searching options (from left to right):
- channel that is used for searching
;
- desired signal front
,
and
- for any;
- signal crossing level (similar to synchronization level) .
fig. 16. View of the searching control item for logic analyser data.
24
By analogy with the synchronization elements for the login analyzer mode the
given item displays search condition mask for all active channels. The searching
control item activates search parameters control panel.
3.7.3 Search parameters control panel
View of the search parameters control panels for two types of data is shown on
fig. 17 and 18.
fig. 17. Search parameters control panel for analogue data.
Panel has the following control items (from left to right):
- channel is used for searching
;
- desired signal front:
- raise,
- fall,
- any;
- signal crossing level
;
- screen positioning variants for searching result:
- on the left,
- in the center,
- on the right;
- run forward direction searching
;
- run backward direction searching .
fig. 18. Search parameters control panel for logic analyser data.
The left side of this panel contains control items, which help to set the search
condition mask for all active channels. It is possible to define any fronts (
,
,
)
and levels ( or
) combinations. It is possible to use different front conditions for
different channels at the same time. If the channel condition is not important then
option
is used. Other control items are identical to the described above for
analogue data search.
In addition to buttons
/
search process can be initiated with the help of
“Operations/Search forward for signal condition” or “Operations/Search backward
for signal condition” menu options, Ctrl+F or Shift+F hot keys combination for
forward and backward searching accordingly.
3.7.4 Selection of oscillogram fragment
Selection of oscillogram fragment is required for a set of different operations,
and is performed with the following steps. Firstly set the mouse cursor so that it
points at the selected fragment beginning or end position and press the left mouse
button. Then the move the mouse cursor in the direction to the end or beginning
position of the selected fragment accordingly, keeping the left mouse button
pressed. Release the left mouse button to finish fragment selection operation. If
the selected fragment is out of the single frame range, frame will be scrolled
25
automatically when the mouse reaches the area close to the screen left or right
borders. It is possible to use “Page Up”, “Page Down”, “Home” and “End” keys
during selection operation to scroll the screen in the required direction.
3.7.5 Horizontal auto-zooming
This function helps to evaluate the optimal zoom value for the selected
oscillogram fragment, in order to place it in the center of the current screen size.
The required oscillogram fragment must be selected before calling this function
(see
3.7.4). The function is run with the help of the toolbar button
,
“View/Selected area auto-zoom” menu option or Alt+Z hot keys combination.
3.7.6 Saving of oscillogram fragment
The selected oscillogram fragment (see 3.7.4) can be saved as a separate file
with the help of this function. The function is run with the help of the toolbar button
or “File/Save selection” menu option. It is more appropriate to save informative
parts of oscillogram than to save large oscillogram files instead.
3.8 Measurement tools
The general measurement items are the measurement panel and a pair of
measurement markers A and B shown on fig. 19.
fig. 19. Measurement panel and A and B markers.
The A and B markers ([4] and [5] accordingly to fig.19) help to measure signal
parameters in the dots, where markers lines cross a signal curve. These
parameters are displayed in the corresponded items of the measurement panel,
for each active channel.
26
3.8.1 Measurement panel
The measurement panel has values displays for each active channel. At the top
of each values display there is a header, see [3] fig.19, showing the channel
number and name. The values display body shows two values: the first one([6]
fig.19) is the signal value for a dot where the line of marker A crosses a signal
curve, the second one([7] fig.19) is the difference value between two values for
dots where lines of marker A and B cross a signal curve. The last item of the
measurement panel is A-B interval display. This item displays the time interval
value between A and B markers, and its corresponding frequency. In the specter
analyzer mode this item displays the frequency for marker A position and
difference of frequencies between A and B markers positions in the horizontal
frequency scale.
The display settings popup menu can be shown by pressing the display header
with the left button of the mouse. This menu contains options which help to display
signal values as voltage, to recalculate them according to the channel name or to
define the unit type recalculation directly from the presented list of the unit types.
The concept of unit types will be considered later see 3.8.4. The values display for
each channel can be hidden or shown again optionally see 3.3.4, the popup menu
option “Hide channel values” has the identical function. It is possible to move the
channel value display to the separate window with the help of “Fly channel display”
popup menu option.
Measurement panel can be attached to the screen body, be displayed in a
separate window or can be the part of the control panel. In the first two cases, the
size of the measurement panel can be changed with the mouse cursor. The size of
the channel value displays is changed according to the measurement panel size.
Menu option “View/Measuring Panel” controls the measurement panel displaying
in the screen body. Menu option “View/Measuring window in control panel” is used
to display the measuring window as a part of the control panel. In this case, the
size of the measuring window depends on the free space available on the control
panel. The last option helps to increase the oscillogram view screen due to more
rational usage of the available space. But it is not always possible due to the
limited space of the control panel with small screen resolutions.
3.8.2 Measurement markers
The measurement markers A and B serve for the signals amplitude and time
parameters measuring, as it was described above. There are several ways to
move measurement markers: with the help of the control panel items [1], [2] fig.19,
with the mouse cursor or keyboard. The non-fixed marker moves within the range
of the screen, and stays on its place during the screen horizontal scrolling. The
fixed marker is bound to a specific oscillogram dot, and moves with it during the
screen scrolling. The cross on the marker handle is a sign of the marker fixation.
As it is shown on fig. 19 marker A [4] is not fixed, marker B [5] is fixed. When
markers are controlled by the control panel elements
[1],
[2] fig.19 their fix
condition can be changed by pressing “Enter” keyboard key. To change the fix
condition with the help of the mouse, move the mouse cursor to the squire that
displays the marker fix condition and press the left mouse button. There is a
27
possibility to move markers by keyboard keys. Press and hold “Shift” or “Ctrl”
keyboard keys to move markers A or B accordingly. The direction of moving is
defined by the left or right arrow keyboard keys. It is possible to use the same
functionality with “Shift” or “Ctrl” keyboard keys to catch markers A or B with the
mouse cursor. Press the left mouse button while “Shift” or “Ctrl” keys are held, and
the marker will move to the current position of the mouse cursor. Release “Shift” or
“Ctrl” keyboard keys and move the marker to the required position, keeping the left
mouse button pressed.
3.8.3 Signal phase and off-duty factor measuring
The program is supplied with the tool that helps with measuring of signal phase
and off-duty factor. The control panel item shown on fig 20 [1], helps with this type
of measurement.
fig. 20. Phase measurement.
The phase or off-duty factor value measuring process starts from the interval
definition. This interval should be the part or a multiple of the signal period, and is
defined with the help of A and B markers. Then the phase/off-duty factor
measurement panel is activates by item [1] fig. 20 of the control panel.
fig. 21. Phase/off-duty factor measurement panel
The first item of this panel fig. 21 helps to define the type of period, defined by
markers A and B such as:
90, 180, 360, 720 degrees - for phase measurement;
100% - defines the whole period for off-duty factor measurement;
“----“
- off phase/off-duty factor measurement.
The next item of this panel shows the value of the full measurement scale. The
program displays the phase/off-duty factor measurement scale ([2] fig. 20) when
the period type is chosen. This scale stays attached to the interval, defined by A
and B markers. The phase/off-duty factor value for any point of the oscillogram can
be found by moving the marker A in relation to the defined scale. This value is
28
displayed as the first value of the item
[1] fig.
20. The second value is the
phase/off-duty factor difference value, according to the A-B interval. According to
fig. 20 the impulse phase on channel 2 is shifted on -4.74 degree relative to the
peak on channel 1, and the phase difference between markers A and B is 175.26
degrees. Select the option of the phase/off-duty factor control item to switch off
phase/off-duty factor measurement.
3.8.4 Signal values recalculation and units types
The program is supplied with a tool that allows signal values recalculation into
values, specified by units types with supplied rules. The unit type term means a set
of parameters and rules description for the value recalculation. Descriptions for all
units types are kept in external files and can be changed or updated by the user.
Each type has a unique name that is used for the definition of the measurement
panel values recalculation. As it was described in 3.8.1, unit type name can be
chosen from the list of measurement panel channel display popup menu. By
default, the unit type name is selected automatically, according to the channel
name. In other words, if the defined channel name is identical to some unit type
name, the Program will automatically use this type for the channel values
recalculation. The unit type name can be chosen as a channel name with the
button of the channel name control item on the channel control panel (see.
3.3.4).
3.8.5 User units types editing
The changing, addition or deletion of the User units types is performed with the
help of “Units types setup” dialog fig. 22. This Dialog is activated by “Control/Units
types setup” menu item or Ctrl+V hot keys combination.
fig. 22. “Units types setup” dialog.
According to figure 22, the dialog has a list of all units types defined by the
user. The left column of this list displays the unit type name, the right one - the
29
corresponding unit. The
“@” prefix symbol will be replaced during the value
formatting with n/u/m/k/m symbols for nano/micro/milli/kilo/mega according to the
current value. If the “@” prefix symbol is not defined the value is displayed as the
absolute value of the current unit. “Formula” field helps to define the formula, used
for the value recalculation according to the current unit type. The formula input
value is defined as “x” symbol. The program supports the next functions:
Function
Comment
abs(x)
Returns the absolute (always positive) value of x
acos(x)
Returns the ar cos function value of x in radians
asin(x)
Returns the ar sin function value of x in radians
atan(x)
Returns the ar tan function value of x in radians
cos(x)
Returns the cos function value of x
exp(x)
Returns the value of exponential function ( ex )
log(x)
Returns the value of natural logarithmic function of x
pow(x, y)
Returns the value of function xy
round(x)
Rounds the value of x to integer
sin(x)
Returns the sin function value of x
sqrt(x)
Returns the squire root of x
tan(x)
Returns the tan function value of x
The next group of items is optional, and defines valid “x” values range for the
current formula. If the current “x” value is out of the range, the value calculation is
ignored, and the result is displayed as “…”.
Changing of the unit type name or unit values can be done with the double click
of the mouse left button on the corresponding list item. Press “Enter” key to accept
the entered value, or “Esc” key to cancel any changes. Any unit type recording is
considered as finished if it has the valid name and formula values. The “Add”
button helps to add the new item type record, “Delete” button is used to delete the
active one. The “Ok” button is used to save all changes made by the user,
“Cancel” button helps to cancel any changes.
30
3.9 User preset files
User preset files (UP) are used to run and configure Device and Program
modules by previously saved preset file. These files keep Device and Program
settings, made by the user for some tasks. UP files also keep plug-ins settings.
3.9.1 Saving of current settings in UP file
The current Device and Program settings can be saved with the help of the tool
bar button
,
“Control/Save user preset” menu item or Alt+R hot keys
combination. When calling this function, the program displays save user preset file
dialog in Fig. 23.
fig. 23. Save user preset file dialog.
The UP file name is the main parameter of this dialog. Parameter “Located in”
helps to define the file location folder. This folder name can be entered to create a
new folder, or it can be selected from the list of existed ones. The “Additional
properties” list helps to define a set of parameters that will be applied in the mode
initiated by the current preset file. When save preset function is executed for the
mode that has been initiated by the UP file, save UP dialog displays the name,
location and properties of this UP file. In this case the user can add corrections in
the current file or save a new one by changing name/location parameters.
3.9.2 Loading or deleting UP files
UP files loading or deleting functions are controlled by the corresponded
elements of the control panel. Functions of these control items show UP files
dropping list. This list contains UP files
or folders that contain UP files. Select
the required file and activate the loading or deleting function with the left mouse
button click, right arrow or “Enter” keyboard keys. Use the left arrow or “Esc”
keyboard keys to move in the root list folder or for function canceling.
31
3.10 Marking of oscillogram
The oscillogram marking function is based on the usage of the bookmarks
functionality. Each bookmark is related to a single oscillogram point and can be
supplied with a comment. The bookmark list is stored in the oscillogram file. That
will help with the search of the oscillogram characteristic points when the file is
opened again. Bookmark markers can be set manually or automatically, when the
analyzer scripts are used (see 3.16). The bookmarks can be set manually while file
viewing or recording.
3.10.1 Bookmark set or delete functions
The toolbar button
, “Bookmark/Set/Remove” menu option or Ctrl+F2 hot keys
combination can be used to set or delete the bookmark. The bookmark position is
defined by the marker A position. When the current marker A position has the
bookmark it will be deleted during the bookmark set/reset function. Use
“Bookmark/Delete all” menu option to delete all bookmarks. Alternatively one or all
bookmarks can be deleted with the help of the bookmarks list window see 3.10.2.
3.10.2 Bookmarks list window
The bookmarks list window is displayed at the bottom of the Program screen.
This window is displayed automatically when the oscillogram file, containing the
bookmarks, is opened. The bookmarks list window can be shown or hidden by the
toolbar button
,
“Bookmark/Show list” menu option or Alt+F2 hot keys
combination. The size of this window can be changed by moving the screen
splitter with the mouse cursor. Activation of any item in the bookmark list moves
the screen position to the position specified by corresponding bookmark.
Oscillogram screen active bookmark changing is caused by the corresponding
item activation in the bookmarks list window. The bookmarks list item can be
changed by the mouse cursor or keyboard up/down arrow keys. The bookmarks
list popup menu can be used to execute some bookmark functions. This menu can
be activated with the mouse right button click inside the list window. The “Delete”
and “Change comment” options of this menu perform corresponding operations
with the list active item. The “Find marker” and “Delete all” popup menu options
serve to find the marker by its comment and clear bookmark list accordingly. The
marker comment can be changed alternatively by the double click of the left
mouse button on the comment field of the required list item. Use “Enter” key to
accept comment entering/changing, or “Esc” key to cancel it.
3.10.3 Navigation through oscillogram with the usage of bookmarks and
search by comment
Oscillogram characteristic points navigation can be realized in two ways. The
first one is using of bookmark list see 3.10.2. Using of toolbar buttons:
and
,
“Bookmark/Move to previous” and “Bookmark/Move to next” or Shift+F2 and F2
keyboard hot keys is the second one. To find the marker by its comment
“Bookmark/Find” menu option can be used.
32
3.10.4 Bookmark play function
During bookmark play function, Program moves oscillogram from one
bookmark to another sequentially, and displays its comment in the popup window.
This function is run by the “Bookmark/Play” menu option. Play function stops when
the last marker is shown. The function can be stopped by using the “Esc” keyboard
key.
3.11 Additional tools
This division describes a set of functions, used in different Program modes.
3.11.1 Using of level/range markers
The level/range markers are used to display a signal level or a range graphically
for each channel separately. These markers are available in real time signal
monitoring and oscillogram file view Program modes. Parameters of level/range
markers are stored in the oscillogram and UP file types. To set, change or delete
level/range marker, toolbar button
,
“Operations/Set level (range) marker” menu
option or Ctrl+L hot keys combination should be used. Execution of this function
shows level/range marker parameters dialog fig. 24.
fig. 24. The level/range marker parameters dialog.
The “For channel” dialog item (fig. 24) allows defining a number of the channel
which level/range marker will be accessed. The next group of items displays
general parameters of the level/range marker for the current channel. The level
marker is defined by “Minimal level” field of this dialog, “Delta level” parameter
must be equal to zero. The range marker is defined by two values: “Minimal level”
and “Delta level” or “Maximal level” optionally (see fig. 24). The level/range marker
comment is the optional parameter, and can be defined by the “Comment” field of
this dialog. Use “OK” button to accept all made modifications, or “Cancel” button to
cancel them. Select the channel number and press “Delete” button to delete the
marker for the chosen channel. Comments of the level/range markers, stored in
the oscillogram file, are shown in the popup window when this oscillogram file is
loaded. Use the close button placed in the right top corner of these windows to
33
close unnecessary comments. The “View/Voltage marker(s) comment” menu item
or Shift+L hot keys combination can be used to show all level/range markers
comment popup windows. The level of transparency for level/range markers is
controlled by the function described in 3.12.3.
3.11.2 Oscillogram file description
There is a possibility to add the oscillogram file description text if needed. This
description will be shown each time when the file is opened. The “View/File
description” menu option can be used to change the file description. This function
execution displays file description dialog window.
3.11.3 View options
Program view options are grouped in the popup menu “View/View Options”. The
following options are defined in the current Program version. The
“White
background” option allows setting the oscillogram screen background color to
white. The “Show vertical scroll bar” option allows switching on/off the vertical
screen scrolling option.
3.12 Screen memory
Usage of screen memory helps to store (accumulate) the current signal shape
into the screen memory. This function is available in all program modes.
3.12.1 Current screen remember function
The screen remember function is initiated by the toolbar button
or
“Operations/Remember screen” menu option. It is possible to do continuous
screen capturing, during real time signals monitoring. This function is activated by
the “Operations/Continuously screen capture” menu option. The program screen
memory accumulates the input signal changes mask, when this function is active.
The same functions are used to stop this process. The screen memory is cleared
automatically when channel view parameters, current mode or time division value
are changed, during continues storage operation.
3.12.2 Saving, loading and clearing of the screen memory
The screen memory can be saved or loaded to or from the image file. These
functions are initiated with the corresponding menu options
“Operations/Save
screen memory”, “Operations/Load screen memory” or “Operations/Clear screen
memory”. Screen memory save function adds a scale and the minimal required
description of channel(s) parameters to the result graphical file.
3.12.3 Changing of screen memory transparency
The transparency value is used to get a better visualization effect for signals
displaying on screen memory content background. The same value is used for
level/range marker displaying. Extension (button at right side) of toolbar button
or “Operations/Transparency level” menu item can be used to change the
34
transparency value. The vertical bar moving of the transparency control window
changes the transparency value.
3.13 Oscillogram printing
The toolbar button
,
“File/Print” menu option or Ctrl+P hot keys combination
should be used to initiate the oscillogram printing process. The oscillogram print
settings dialog fig. 25 is shown, when the print function is executed.
fig. 25. Oscillogram print settings dialog.
Print configuration starts with the selection of the printer which will be used for
printing. Then choose the required “Print range” value from the following values:
“Current screen”, “Selection” or “All”. According to this option the Program will print
current screen data, the selected part of the oscillogram or the whole oscillogram
file. Be careful with using the last option, this can cause printing of the big amount
of pages. The “Preview” dialog button activates the preview window (see 3.13.1).
The “
” button is used to start printing process, “Cancel” button cancels printing.
35
3.13.1 Print preview window
The print preview window fig. 26 shows the graphic information that will be
printed and allows the user to adjust it if necessary.
fig. 26. Print preview window.
The control panel, placed in the top of preview window, helps to control the
page number and the current viewing zoom. The control panel of the signals
parameters is placed in the left side of this window. This panel allows setting the
optimal printing parameters for each channel signal. This panel has the same
control items as the channel parameters control panel see 3.3.4.
3.14 Saving of oscillogram fragment in image file
This function allows saving the interesting fragment of the oscillogram in the
image file, supplied with the required information of channels parameters.
3.14.1 Selecting of oscillogram area
This function allows selecting the required size of the rectangular area in the
horizontal and vertical space of the oscillogram. The User defines horizontal and
vertical sizes of the resulting image to optimize the view of signals in the defined
rectangular space. It should be noted that the Program will increase the vertical
size of the resulting image to add the required information. The area select
function can be started from the horizontal fragment selection described in 3.7.4.
36
Then Program switches into the rectangular area selecting mode by pressing the
toolbar button . When the function is started with the usage of the toolbar button
directly (without horizontal fragment selection), the Program selects the area
equivalent to the current screen size. Rectangular area size can be changed by
simple moving of area borders with the mouse cursor. The popup window in this
case displays sizes of the current area in dots.
3.14.2 Saving oscillogram fragment as a picture
In order to save the oscillogram fragment into the graphic file, it is necessary to
select the rectangular area (see 3.14.1). The function is run with the help of the
toolbar button or “File/Save as picture” menu item.
3.15 Plug-ins
Plug-in term is used for external embedded modules, allowing to analyze the
Device data stream in a real time, and to perform the signals visualization and
parameters measuring. Using of plug-ins allows extending the Program
functionality for real time specific signals processing. It is possible to create and
use customer modules. The description of plug-in modules structure is not
included into this document.
3.15.1 Using of Plug-ins
Loading of plug-ins modules can be done manually or automatically with UP
files see 3.9.2. There are two ways for manual plug-in loading, depending on its
type and functionality. The first way is when the Device is started in the required
mode, and plug-in is loaded next. The second way is when plug-in is loaded first,
and then plug-in configures and runs the current Device in the required mode. It
should be noted that not all plug-in types can run the Device, therefore the second
way can not be used for all plug-ins modules.
Use “Load plug-in” item of the control panel or list of items in the bottom of the
“Control” menu to load the required plug-in. Plug-in control panel fig. 27 is placed
at the bottom of the screen when plug-in is loaded.
37
fig. 27. Program window when plug-in is loaded.
The Plug-in control panel contains items that control plug-in functionality and
display investigated signal parameters. Some plug-ins can create the graphical
panel (placed in the right side on fig 27), that displays data, processed by plug-in
as graphics, diagrams and etc. The size of this panel can be changed with the help
of the splitter moving to the left or right with the mouse cursor. As it is shown on
fig.27, plug-in graphic panel can contain separate items. Each item has a header
and a minimize button placed in the right side of it. This button can be used to hide
or restore item context displaying. When the main part of the item header is
pressed with the mouse left button, its context menu is displayed. When the plug-
in module is loaded, some Program control items can be unavailable, depending
on the current plug-in functionality.
38
3.16 Oscillogram data automatic analysis
Ability of the oscillogram automatic analysis is based on the integration of
Program objects with the operation system script core. Program allows using a
general set of JScript (Java Script) and VBScript (Visual Basic Script) functions
and objects, and adds a set of own ones. Program's objects, integrated in the
script core, allow accessing and analyzing of the oscillogram data, displaying
results of this analysis in the text or graphical view, marking and commenting
oscillogram specific points.
The script files are text or HTML files, which contain instructions for data
processing and result visualization. In addition the Program is supplied with the
encoded file type, used for source algorithm protection purpose. The current
Program version supports two types of analysis. The first one is used for the
analysis of the whole or selected part of the oscillogram, with receiving of all types
of results and oscillogram marking. This type is based on the analyzer script files.
The second one is the functional extension of the Program interface, which
extends Program abilities for a specific signal processing while oscillogram file
viewing. This type is based on the analyzer panel files. As well as plug-ins,
analyser panels are integrated into the Program interface and have own control
items, placed on a separate panel at the bottom of the screen. Program script
objects and functions description are not included into this document.
3.16.1 Using of analyser scripts
Analyser script files are files with extension “*.ajs” - containing Jscript source,
“*.abs”
- for VBScript source or HTML files. The “*.ab ” extension file is the
encoded file type which can contain source of all described above types. The
toolbar button
or “Analysis/Open script” menu option is used to load the script
file. The toolbar button
,
“Analysis/Execute script” menu option or Ctrl+A hot
keys combination is used to run the script. The toolbar button or “Analysis/Abort
script” menu option is used to cancel the script execution. Toolbar buttons
,
or
“View/Switch to oscillogram view”, “View/Switch to report view” menu items or
Alt+W, Alt+R hot keys combination can be used to switch oscillogram and report
view modes accordingly. Use toolbar button
or “Analysis/Delete report” menu
option to delete results of analysis. When the script creates several elements,
placed in the report window, the tab control item on the top of the report window or
Ctrl+Tab hot keys combination serves to switch active ones. The name of the
current script file can be attached to the oscillogram file. In this case, the attached
script file will be loaded each time when the oscillogram file is loaded. The
“Analysis/Attach current script” menu option can be used to attach the loaded
script file. To cancel any attached script use the “Analysis/Detach script” menu
option. The
“Analysis/Run after load” menu option activates script execution
directly after loading.
3.16.2 Analyzer panel usage
Analyzer panel files have “*.apn” file extension, and “*.ap ” for encoded analyzer
panel files. The toolbar button or “Analysis/Load analyser panel” menu option is
39
used to load the analyzer panel file. The running of the analyzer panel file is
performed directly after the file loading. The Analyser panel closing stops its script
execution. The name of the current analyzer panel can be attached to the
oscillogram file with the “Analysis/Attach current panel” menu option. In this case
the panel is loaded automatically when such oscillogram file is loaded. To cancel
this attachment use the “Analysis/Detach panel” menu option.
3.16.3 Graphical results printing
This function serves to receive a printing copy of the graphics drawn as an
oscillogram data analyze result. Printing function is initiated by the toolbar button
,
“File/Print” menu option or Ctrl+P hot keys combination. The graphics print
setup dialog fig. 28 helps to configure the print function.
fig. 28. Graphics print setup dialog.
Print configuration starts with the selection of the printer which will be used for
printing. Then the required “Print range” value should be chosen from the following
values “Current screen” or “Whole graphic”. According to this option the Program
will print the current screen view or the whole graphic. The “Preview” dialog button
activates the preview window. The
” button is used to start the printing
process, “Cancel” button cancels printing.
40
3.16.4 Saving of graphic in a picture file
This function helps to save graphical results of oscillogram data analysis, as a
picture file. The function is initiated with the toolbar button
or “File/Save as
picture” menu option. The “Save as picture” dialog fig. 29 is used to configure
function options.
fig. 29. The “Save as picture” dialog.
As it is shown on fig. 29, the dialog has control items that serve to define the
result image size, select what will be displayed in the result image “Current screen”
or “Whole data” - for the whole graphic. The “Preview” dialog button activates the
preview window. The “Save” button is used to save the picture file, “Cancel” button
cancels it.
41
3.17 Emulation mode
This mode allows emulating the real time signal monitoring on the oscillogram
file, previously stored without the real Device usage. This mode has several
limitations in comparison with the real process. In spite of this it allows using the
whole synchronization, measurement and plug-ins tools set, with the imitation of
real signals.
3.17.1 Running of emulation
The required oscillogram file, which data will be used for emulation, should be
loaded firstly before emulation mode can be started. As well as the real Device,
the emulator is run by the “Start device” item of the control panel. The “Program
emulator” item should be chosen as the data source Device fig. 30.
fig. 30. Start device panel with the “Program emulator” device.
The device mode and channels number items of this panel are inactive, they
display the mode and number of channels in data source file, used for the
emulation. The next item is used to switch on or off data circling option. This
option allows circling of emulation mode when the emulator reaches the end of
emulation source file.
3.17.2 Emulation activation
This option sets the emulator device as a default device. The “Control/Activate
emulation” menu item is used to activate the emulation priority. It is necessary to
use this option when the emulation mode should be used with UP-files or plug-ins
loading. Otherwise, if this option is disabled, the Program will make attempts to run
any available Device.
42
Open full version in PDF - 1 MB